CVE-2024-48966

Life2000 service tools for test and calibration do not support user authentication

The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

CVSS 3.1 · 10 Critical EPSS < 1% KEV · Not listed
Vendors
CVE-2024-9834

Improper data protection on Life2000 ventilator serial interface

Improper data protection on the ventilator's serial interface could allow an attacker to send and receive messages that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

CVSS 3.1 · 9.3 Critical EPSS < 1% KEV · Not listed
Vendors
CVE-2024-9832

No limit on failed login attempts with Clinician Password or Serial Number Clinician Password on Life2000 Ventilator

There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.

CVSS 3.1 · 9.3 Critical EPSS < 1% KEV · Not listed
Vendors
CVE-2024-48974

Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates

The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.

CVSS 3.1 · 9.3 Critical EPSS < 1% KEV · Not listed
Vendors
CVE-2024-48973

Debug port on Life2000 Ventilator serial interface is enabled by default

The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

CVSS 3.1 · 9.3 Critical EPSS < 1% KEV · Not listed
Vendors
CVE-2024-48971

Clinician Password and Serial Number Clinician Password are hard-coded in Life2000 Ventilator

The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privileges.

CVSS 3.1 · 9.3 Critical EPSS < 1% KEV · Not listed
Vendors
CVE-2024-48970

Life2000 Ventilator microcontroller lacks memory protection

The ventilator's microcontroller lacks memory protection. An attacker could connect to the internal JTAG interface and read or write to flash memory using an off-the-shelf debugging tool, which could disrupt the function of the device and/or cause unauthorized information disclosure.

CVSS 3.1 · 9.3 Critical EPSS < 1% KEV · Not listed
Vendors
CVE-2024-48967

Life2000 ventilator and Service PC lack sufficient audit logging capabilities

The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance.

CVSS 3.1 · 10 Critical EPSS < 1% KEV · Not listed
Vendors
CVE-2022-26392

Format String vulnerability

The Baxter Spectrum WBM (v16, v16D38) and Baxter Spectrum WBM (v17, v17D19, v20D29 to v20D32) when in superuser mode is susceptible to format string attacks via application messaging. An attacker could use this to read memory in the WBM to access sensitive information.

CVSS 3.1 · 3.1 Low EPSS < 1% KEV · Not listed
Vendors
CVE-2022-26390

Unencrypted internal storage of security credentials

The Baxter Spectrum Wireless Battery Module (WBM) stores network credentials and PHI (only applicable to Spectrum IQ pumps using auto programming) in unencrypted form. An attacker with physical access to a device that hasn't had all data and settings erased may be able to extract sensitive information.

CVSS 3.1 · 4.2 Medium EPSS < 1% KEV · Not listed
Vendors
CVE-2021-43935

ICSMA-21-343-01 Hillrom Welch Allyn Cardio Products

The impacted products, when configured to use SSO, are affected by an improper authentication vulnerability. This vulnerability allows the application to accept manual entry of any active directory (AD) account provisioned in the application without supplying a password, resulting in access to the application as the supplied AD account, with all associated privileges.

CVSS 3.1 · 8.1 High EPSS 1.1% KEV · Not listed
Vendors
CVE-2020-12048

Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmitting treatment and prescription data on the network between the Phoenix system and the Exalis dialysis data management tool. An attacker with access to the network could observe sensitive treatment and prescription data sent between the Phoenix system and the Exalis tool.

CVSS 3.1 · 7.5 High EPSS < 1% KEV · Not listed
Vendors
CVE-2020-12041

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits temporary configuration changes to network settings of the WBM, and allows the WBM to be rebooted. Temporary configuration changes to network settings are removed upon reboot.

CVSS 3.1 · 9.4 Critical EPSS 1.4% KEV · Not listed
Vendors
CVE-2020-12040

Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.

CVSS 3.1 · 9.8 Critical EPSS < 1% KEV · Not listed
Vendors
CVE-2020-12039

Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hardcoded passwords when physically entered on the keypad provide access to biomedical menus including device settings, view calibration values, network configuration of Sigma Spectrum WBM if installed.

CVSS 3.1 · 2.4 Low EPSS < 1% KEV · Not listed
Vendors