Life2000 service tools for test and calibration do not support user authentication
Default status is the baseline for the product, each version can override it (e.g. patched versions marked unaffected).
| Vendor | Product | Default status | Versions | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Baxter | Life2000 Ventilation System | unaffected |
|
No data.
No data.
No data available yet.
| Source | ID | Title |
|---|---|---|
EUVD |
EUVD-2024-43153 | The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance. |
No CVSS v4.0
No CVSS v3.0
No CVSS v2
This CVE is not in the KEV list.
Exploit prediction score: < 1% (0.695% probability within 30 days)
| Type | Values Removed | Values Added |
|---|---|---|
| First Time appeared |
Baxter
Baxter life2000 Ventilator Firmware |
|
| CPEs | cpe:2.3:o:baxter:life2000_ventilator_firmware:*:*:*:*:*:*:*:* | |
| Vendors & Products |
Baxter
Baxter life2000 Ventilator Firmware |
|
| Metrics |
ssvc
|
| Type | Values Removed | Values Added |
|---|---|---|
| Description | The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance. | |
| Title | Life2000 service tools for test and calibration do not support user authentication | |
| Weaknesses | CWE-306 | |
| References |
| |
| Metrics |
cvssV3_1
|