CVE-2024-48966

Life2000 service tools for test and calibration do not support user authentication

Published Nov 14, 2024 CVSS 3.1 · 10 Critical EPSS < 1% KEV · Not listed
Description
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

Remediation

No solution or workaround provided in the CVE record

Tracking

Sign in to view the affected projects.

Advisories

Source ID Title
EUVD EUVD EUVD-2024-43153 The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
History
Type Values Removed Values Added
First Time appeared Baxter
Baxter life2000 Ventilator Firmware
CPEs cpe:2.3:o:baxter:life2000_ventilator_firmware:*:*:*:*:*:*:*:*
Vendors & Products Baxter
Baxter life2000 Ventilator Firmware
Metrics ssvc

{'options': {'Automatable': 'yes', 'Exploitation': 'none', 'Technical Impact': 'total'}, 'version': '2.0.3'}

Type Values Removed Values Added
Description The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
Title Life2000 service tools for test and calibration do not support user authentication
Weaknesses CWE-306
References
Metrics cvssV3_1

{'score': 10, 'vector': 'CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H'}