CVE-2018-8870

Medtronic MyCareLink Patient Monitor Use of Hard-coded Password

Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains a hard-coded operating system password. An attacker with physical access can remove the case of the device, connect to the debug port, and use the password to gain privileged access to the operating system.

CVSS 3.1 · 6.4 Medium EPSS < 1% KEV · Not listed
Vendors
CVE-2018-8868

Medtronic MyCareLink Patient Monitor Exposed Dangerous Method or Function

Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.

CVSS 3.1 · 6.2 Medium EPSS < 1% KEV · Not listed
Vendors