Medtronic MyCareLink Patient Monitor Exposed Dangerous Method or Function
Default status is the baseline for the product, each version can override it (e.g. patched versions marked unaffected).
| Vendor | Product | Default status | Versions | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Medtronic | 24950 MyCareLink Monitor | unaffected |
|
||||||
| Medtronic | 24952 MyCareLink Monitor | unaffected |
|
Configuration 1 [-]
| AND |
|
Configuration 2 [-]
| AND |
|
No data.
No data available yet.
Vendor Solution
Medtronic will release several rolling over-the-air product updates that will mitigate the vulnerabilities described within this advisory. These updates will be applied to devices automatically as part of standard, reoccurring update processes. In addition, Medtronic has increased security monitoring of affected devices and related infrastructure. Medtronic has released additional patient focused information, at the following location: https://www.medtronic.com/security
Vendor Workaround
Medtronic recommends users take additional defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should: * Maintain good physical controls over the home monitor as the best mitigation to these vulnerabilities. * Only use home monitors obtained directly from their healthcare provider or a Medtronic representative to ensure integrity of the system. * Report any concerning behavior regarding their home monitor to their healthcare provider or a Medtronic representative. Medtronic has released additional patient focused information, at the following location: https://www.medtronic.com/security
| Source | ID | Title |
|---|---|---|
EUVD |
EUVD-2018-20476 | Medtronic MyCareLink Patient Monitor, 24950 MyCareLink Monitor, all versions, and 24952 MyCareLink Monitor, all versions, contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can apply the other vulnerabilities within this advisory to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality. |
No CVSS v4.0
This CVE is not in the KEV list.
Exploit prediction score: < 1% (0.336% probability within 30 days)
Key SSVC decision points have not yet been added.
| Type | Values Removed | Values Added |
|---|---|---|
| Description | Medtronic MyCareLink Patient Monitor, 24950 MyCareLink Monitor, all versions, and 24952 MyCareLink Monitor, all versions, contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can apply the other vulnerabilities within this advisory to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality. | Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality. |
| Title | Medtronic MyCareLink Patient Monitor Exposed Dangerous Method or Function | |
| References |
| |
| Metrics |
cvssV3_1
|