CVE-2023-25931

Medtronic Micro Clinician & InterStim X Clinician App Password Reset Issue

Published Mar 1, 2023 CVSS 3.1 · 6.4 Medium EPSS < 1% KEV · Not listed
Description
Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.

Remediation

Vendor Solution

Current versions of the application has mitigated this vulnerability. Please refer to the Medtronic Security Bulletin for update guidance. 

Tracking

Sign in to view the affected projects.

Advisories

Source ID Title
EUVD EUVD EUVD-2023-29819 Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.
History
Type Values Removed Values Added
Metrics ssvc

{'options': {'Automatable': 'no', 'Exploitation': 'none', 'Technical Impact': 'partial'}, 'version': '2.0.3'}

Weaknesses

Sources

Status
PUBLISHED
Assigner
Medtronic
Published
Updated
Mar 7, 2025, 9:34 PM
Reserved
Feb 16, 2023, 5:24 PM
Status
Modified
Published
Mar 1, 2023, 7:15 PM
Modified
Jun 17, 2026, 5:42 AM