CVE-2019-13535

Medtronic Valleylab FT10 and LS10 Protection Mechanism Failure

Published Nov 8, 2019 CVSS 3.1 · 4.6 Medium EPSS < 1% KEV · Not listed
Description
In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States) version 1.20.2 and lower, the RFID security mechanism does not apply read protection, allowing for full read access of the RFID security mechanism data.

Remediation

Vendor Solution

A software patch is available now for the affected Valleylab platforms. If you suspect you are in possession of an instrument that is not FDA approved or cleared to be used with Medtronic Valleylab FT10 or LS10, please contact Medtronic or your medical device supplier. If you have concerns about FDA clearance or approval of current or future instruments, please contact your medical device supplier. Please contact https://www.medtronic.com/security

Tracking

Sign in to view the affected projects.

Advisories

Source ID Title
EUVD EUVD EUVD-2019-4990 In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States) version 1.20.2 and lower, the RFID security mechanism does not apply read protection, allowing for full read access of the RFID security mechanism data.
Weaknesses

Sources

Status
PUBLISHED
Assigner
icscert
Published
Updated
May 22, 2025, 6:39 PM
Reserved
Jul 11, 2019, 12:00 AM
Status
Modified
Published
Nov 8, 2019, 8:15 PM
Modified
Jun 17, 2026, 2:16 AM