Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.
Search vulnerabilities
CVE-2020-12037
CVSS 3.1 · 7.5 High
EPSS < 1%
KEV · Not listed
Vendors
Products
CVE-2020-12036
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.
CVSS 3.1 · 7.5 High
EPSS < 1%
KEV · Not listed
Vendors
Products
CVE-2020-12035
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings, calibration settings, and network configuration. This could allow an attacker to modify device settings and calibration.
CVSS 3.1 · 4.9 Medium
EPSS < 1%
KEV · Not listed
Vendors