Default status is the baseline for the product, each version can override it (e.g. patched versions marked unaffected).
| Vendor | Product | Default status | Versions | ||||||
|---|---|---|---|---|---|---|---|---|---|
| n/a | n/a | affected |
|
Configuration 1 [-]
| AND |
|
No data.
| Vendor | Product | Versions |
|---|---|---|
| Thermofisher | Ion Torrent Onetouch 2 | — |
No CVSS v4.0
No CVSS v3.0
No CVSS v2
This CVE is not in the KEV list.
Exploit prediction score: < 1% (0.499% probability within 30 days)
| Type | Values Removed | Values Added |
|---|---|---|
| First Time appeared |
Thermofisher ion Torrent Onetouch 2 Firmware
|
|
| CPEs | cpe:2.3:h:thermofisher:ion_torrent_onetouch_2:-:*:*:*:*:*:*:* cpe:2.3:o:thermofisher:ion_torrent_onetouch_2_firmware:-:*:*:*:*:*:*:* |
|
| Vendors & Products |
Thermofisher ion Torrent Onetouch 2 Firmware
|
| Type | Values Removed | Values Added |
|---|---|---|
| Weaknesses | CWE-521 | |
| Metrics |
cvssV3_1
|
| Type | Values Removed | Values Added |
|---|---|---|
| First Time appeared |
Thermofisher
Thermofisher ion Torrent Onetouch 2 |
|
| Vendors & Products |
Thermofisher
Thermofisher ion Torrent Onetouch 2 |
| Type | Values Removed | Values Added |
|---|---|---|
| Description | An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. They run an SSH server accessible over the default port 22. The root account has a weak default password of ionadmin, and a password change policy for the root account is not enforced. Thus, an attacker with network connectivity can achieve root code execution. NOTE: This vulnerability only affects products that are no longer supported by the maintainer. | |
| References |
|