Philips Intellispace Cardiovascular (ISCV) Use of Weak Credentials
Default status is the baseline for the product, each version can override it (e.g. patched versions marked unaffected).
| Vendor | Product | Default status | Versions | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Philips | Intellispace Cardiovascular (ISCV) | unaffected |
|
No data.
No data.
No data available yet.
Vendor Solution
Philips recommends the following mitigations: * Resolved in ISCV 4.2 build 20589, which was released in May 2019. * Philips recommends users upgrade ISCV installed base to the latest ISCV version (at the time of this publication is 830089 – IntelliSpace Cardiovacular 8.0.0.0) * Please contact a local Philips sales (service) representative to learn how to engage this upgrade process. * For managed services users, new releases will be made available upon resource availability. Releases are subject to country-specific regulations. Refer to the Philips advisory https://www.philips.com/a-w/security/security-advisories.html for more details.
| Source | ID | Title |
|---|---|---|
EUVD |
EUVD-2025-6389 | A token is created using the username, current date/time, and a fixed AES-128 encryption key, which is the same across all installations. |
No CVSS v3.0
No CVSS v2
This CVE is not in the KEV list.
Exploit prediction score: < 1% (0.127% probability within 30 days)
| Type | Values Removed | Values Added |
|---|---|---|
| Metrics |
epss
|
epss
|
| Type | Values Removed | Values Added |
|---|---|---|
| Metrics |
ssvc
|
| Type | Values Removed | Values Added |
|---|---|---|
| Description | A token is created using the username, current date/time, and a fixed AES-128 encryption key, which is the same across all installations. | |
| Title | Philips Intellispace Cardiovascular (ISCV) Use of Weak Credentials | |
| Weaknesses | CWE-1391 | |
| References |
| |
| Metrics |
cvssV3_1
|
No data.