Remote Code Execution Vulnerability in Radiometer Products
Default status is the baseline for the product, each version can override it (e.g. patched versions marked unaffected).
| Vendor | Product | Default status | Versions | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Radiometer Medical Aps | ABL90 FLEX and ABL90 FLEX PLUS Analyzers | unaffected |
|
|||||||||
| Radiometer Medical Aps | ABL800 BASIC and ABL800 FLEX Analyzers | unaffected |
|
|||||||||
| Radiometer Medical Aps | AQT90 FLEX Analyzers | unaffected |
|
No data.
No data.
| Vendor | Product | Versions |
|---|---|---|
| Radiometer | Abl800 Basic Analyzer | — |
| Radiometer | Abl800 Flex Analyzer | — |
| Radiometer | Abl90 Flex Analyzer | — |
| Radiometer | Abl90 Flex Plus Analyzer | — |
| Radiometer | Aqt90 Flex Analyzer | — |
Vendor Solution
Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution.
Vendor Workaround
If the network is not considered secure, please remove the analyzer from the network.
No CVSS v4.0
No CVSS v3.0
No CVSS v2
This CVE is not in the KEV list.
Exploit prediction score: < 1% (0.465% probability within 30 days)
| Link | Providers |
|---|---|
| https://www.radiometer.com/myradiometer |
|
| Type | Values Removed | Values Added |
|---|---|---|
| Metrics |
ssvc
|
| Type | Values Removed | Values Added |
|---|---|---|
| First Time appeared |
Radiometer
Radiometer abl800 Basic Analyzer Radiometer abl800 Flex Analyzer Radiometer abl90 Flex Analyzer Radiometer abl90 Flex Plus Analyzer Radiometer aqt90 Flex Analyzer |
|
| Vendors & Products |
Radiometer
Radiometer abl800 Basic Analyzer Radiometer abl800 Flex Analyzer Radiometer abl90 Flex Analyzer Radiometer abl90 Flex Plus Analyzer Radiometer aqt90 Flex Analyzer |
| Type | Values Removed | Values Added |
|---|---|---|
| Description | A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software. Other related CVE's are CVE-2025-14095 & CVE-2025-14096. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer. Temporary work Around: If the network is not considered secure, please remove the analyzer from the network. Permanent solution: Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication. | |
| Title | Remote Code Execution Vulnerability in Radiometer Products | |
| Weaknesses | CWE-287 | |
| References |
| |
| Metrics |
cvssV3_1
|