CVE-2025-14096

Credential Disclosure vulnerability in Radiometer Products

Published Dec 17, 2025 CVSS 3.1 · 8.4 High EPSS < 1% KEV · Not listed
Description
A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.

Other related CVE's are CVE-2025-14095 & CVE-2025-14097.



Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.





Required Configuration for Exposure:


Attacker requires physical access to the analyzer.



Temporary work Around:
Only authorized people can physically access the analyzer.

Permanent solution:
Local Radiometer representatives will contact all affected customers to discuss a permanent solution.

Exploit Status:


Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.

Remediation

Vendor Solution

Local Radiometer representatives will contact all affected customers to discuss a permanent solution.


Vendor Workaround

Only authorized people can physically access the analyzer.

Tracking

Sign in to view the affected projects.

Advisories

No advisories available
References
History
Type Values Removed Values Added
Metrics ssvc

{'options': {'Automatable': 'no', 'Exploitation': 'none', 'Technical Impact': 'total'}, 'version': '2.0.3'}

Type Values Removed Values Added
First Time appeared Radiometer
Radiometer abl800 Basic Analyzer
Radiometer abl800 Flex Analyzer
Radiometer abl90 Flex Analyzer
Radiometer abl90 Flex Plus Analyzer
Radiometer aqt90 Flex Analyzer
Vendors & Products Radiometer
Radiometer abl800 Basic Analyzer
Radiometer abl800 Flex Analyzer
Radiometer abl90 Flex Analyzer
Radiometer abl90 Flex Plus Analyzer
Radiometer aqt90 Flex Analyzer
Type Values Removed Values Added
Description A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.                                                                                                                                                                                  Other related CVE's are CVE-2025-14095 and CVE-2025-14097.                                                                                               Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.                   Required Configuration for Exposure: Attacker requires physical access to the analyzer.                                                              Temporary work Around: Only authorized people can physically access the analyzer.                                                                      Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution.                                                                                                                                                                                                     Exploit Status: Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication. A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system. Other related CVE's are CVE-2025-14095 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Attacker requires physical access to the analyzer. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.
Type Values Removed Values Added
Description A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.                                                                                                                                                                                  Other related CVE's are CVE-2025-14095 and CVE-2025-14097.                                                                                               Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.                   Required Configuration for Exposure: Attacker requires physical access to the analyzer.                                                              Temporary work Around: Only authorized people can physically access the analyzer.                                                                      Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution.                                                                                                                                                                                                     Exploit Status: Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.
Title Credential Disclosure vulnerability in Radiometer Products
Weaknesses CWE-250
CWE-798
References
Metrics cvssV3_1

{'score': 8.4, 'vector': 'CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H'}

Weaknesses

Sources

Status
PUBLISHED
Assigner
Radiometer
Published
Updated
Dec 17, 2025, 2:43 PM
Reserved
Dec 5, 2025, 10:50 AM
Status
Deferred
Published
Dec 17, 2025, 1:15 PM
Modified
Jun 17, 2026, 8:35 AM