Vendor Workaround
Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks. If asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120.
| Source | ID | Title |
|---|---|---|
EUVD |
EUVD-2025-5402 | The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device. |
No CVSS v3.1
No CVSS v3.0
No CVSS v2
This CVE is not in the KEV list.
Exploit prediction score: < 1% (0.483% probability within 30 days)
| Type | Values Removed | Values Added |
|---|---|---|
| Metrics |
ssvc
|
| Type | Values Removed | Values Added |
|---|---|---|
| Description | The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device. | |
| Weaknesses | CWE-912 | |