Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor
Default status is the baseline for the product, each version can override it (e.g. patched versions marked unaffected).
| Vendor | Product | Default status | Versions | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Contec Health | CMS8000 Patient Monitor | unaffected |
|
No data.
No data.
No data available yet.
Vendor Workaround
Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks. If asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120. Please note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication .
| Source | ID | Title |
|---|---|---|
EUVD |
EUVD-2025-1797 | The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device. |
No CVSS v3.0
No CVSS v2
This CVE is not in the KEV list.
Exploit prediction score: 1.1% (1.141% probability within 30 days)
| Type | Values Removed | Values Added |
|---|---|---|
| Description | Contec Health CMS8000 Patient Monitor sends out remote access requests to a hard-coded IP address, bypassing existing device network settings to do so. This could serve as a backdoor and lead to a malicious actor being able to upload and overwrite files on the device. | The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device. |
| Type | Values Removed | Values Added |
|---|---|---|
| Metrics |
ssvc
|
| Type | Values Removed | Values Added |
|---|---|---|
| Description | The affected product sends out remote access requests to a hard-coded IP address, bypassing existing device network settings to do so. This could serve as a backdoor and lead to a malicious actor being able to upload and overwrite files on the device. | Contec Health CMS8000 Patient Monitor sends out remote access requests to a hard-coded IP address, bypassing existing device network settings to do so. This could serve as a backdoor and lead to a malicious actor being able to upload and overwrite files on the device. |
| Metrics |
cvssV3_1
|
| Type | Values Removed | Values Added |
|---|---|---|
| Description | The affected product sends out remote access requests to a hard-coded IP address, bypassing existing device network settings to do so. This could serve as a backdoor and lead to a malicious actor being able to upload and overwrite files on the device. | |
| Title | Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor | |
| Weaknesses | CWE-912 | |