CVE-2024-1275

Vulnerability in Baxter Welch Allyn Connex Spot Monitor

Published May 31, 2024 CVSS 4.0 · 9.1 Critical EPSS < 1% KEV · Not listed
Description
Use of Default Cryptographic Key vulnerability in Baxter Welch Allyn Connex Spot Monitor may allow Configuration/Environment Manipulation.This issue affects Welch Allyn Connex Spot Monitor in all versions prior to 1.52.

Remediation

Vendor Solution

Baxter has released a software update for all impacted devices and software to address this vulnerability. A new version of the product that mitigates the vulnerability is available as follows: * Welch Allyn Connex Spot Monitor: Version 1.52.01 (available October 16, 2023) Baxter recommends users upgrade to the latest versions of their products. Information on how to update products to their new versions can be found on the Baxter disclosure page https://www.baxter.com/product-security  or the Hillrom disclosure page https://www.hillrom.com/en/responsible-disclosures/ . Baxter recommends the following workarounds to help reduce risk: * Apply proper network and physical security controls. * Ensure a unique encryption key is configured and applied to the product (as described in the Connex Spot Monitor Service Manual).

Tracking

Sign in to view the affected projects.

Advisories

Source ID Title
EUVD EUVD EUVD-2024-17036 Use of Default Cryptographic Key vulnerability in Baxter Welch Allyn Connex Spot Monitor may allow Configuration/Environment Manipulation.This issue affects Welch Allyn Connex Spot Monitor in all versions prior to 1.52.
History
Type Values Removed Values Added
Metrics epss

{'score': 0.00115}

epss

{'score': 0.00192}

Subscriptions

No data.

Weaknesses

Sources

Status
PUBLISHED
Assigner
Baxter
Published
Updated
Aug 1, 2024, 6:33 PM
Reserved
Feb 6, 2024, 2:20 PM
Status
Deferred
Published
May 31, 2024, 6:15 PM
Modified