Philips Vue PACS Use of Default Credentials
Default status is the baseline for the product, each version can override it (e.g. patched versions marked unaffected).
| Vendor | Product | Default status | Versions | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Philips | Vue PACS | unaffected |
|
No data.
No data available yet.
Vendor Solution
Philips recommends the following mitigations: * For CVE-2021-28165, Philips recommends configuring the Vue PACS environment per D000763414 – Vue_PACS_12_Ports_Protocols_Services_Guide available on Incenter https://incenter.medical.philips.com/Default.aspx . Vue PACS version 12.2.8.410* released in October 2023 prevents this vulnerability. * For CVE-2023-40704, Philips recommends no action needed due to low risk of exploitability, but customers can request that Philips update database password(s). For managed services users, new releases will be made available upon resource availability. Releases are subject to country specific regulations. Users with questions regarding their specific Philips Vue PACS installations and new release eligibility should contact their local Philips sales representative or submit a request in the Philips Informatics Support portal https://www.informatics.support.philips.com/csm . Refer to the Philips advisory https://www.philips.com/productsecurity for more details.
| Source | ID | Title |
|---|---|---|
EUVD |
EUVD-2023-45259 | The product does not require unique and complex passwords to be created during installation. Using Philips's default password could jeopardize the PACS system if the password was hacked or leaked. An attacker could gain access to the database impacting system availability and data integrity. |
No CVSS v3.0
No CVSS v2
This CVE is not in the KEV list.
Exploit prediction score: < 1% (0.338% probability within 30 days)
| Type | Values Removed | Values Added |
|---|---|---|
| Description | Philips Vue PACS uses default credentials for potentially critical functionality. | The product does not require unique and complex passwords to be created during installation. Using Philips's default password could jeopardize the PACS system if the password was hacked or leaked. An attacker could gain access to the database impacting system availability and data integrity. |
| Metrics |
cvssV3_1
|
cvssV3_1
|
| Type | Values Removed | Values Added |
|---|---|---|
| Weaknesses | NVD-CWE-Other |