Authenticated eval injection in B. Braun Space Battery pack SP with Wi-Fi
Default status is the baseline for the product, each version can override it (e.g. patched versions marked unaffected).
| Vendor | Product | Default status | Versions | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| B. Braun | Space Battery Pack SP with Wi-Fi | unaffected |
|
Configuration 1 [-]
| AND |
|
No data.
No data available yet.
Vendor Solution
Update to version L93/U93 or newer solves the issue.
Vendor Workaround
Use secure credentials and keep them secret. Secure your network to prevent outside access to medical devices.
| Source | ID | Title |
|---|---|---|
EUVD |
EUVD-2023-12877 | An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks |
No CVSS v4.0
No CVSS v3.0
No CVSS v2
This CVE is not in the KEV list.
Exploit prediction score: < 1% (0.604% probability within 30 days)
| Type | Values Removed | Values Added |
|---|---|---|
| Metrics |
ssvc
|