Contec Health CMS8000
Default status is the baseline for the product, each version can override it (e.g. patched versions marked unaffected).
| Vendor | Product | Default status | Versions | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Contec Health | CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor | affected |
|
Configuration 1 [-]
| AND |
|
No data.
No data available yet.
Vendor Workaround
Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information. The following mitigations could assist in reducing the risk for exploitation of vulnerabilities: Disabling UART functionality at the CPU level Enforcing unique device authentication before granting access to the terminal / bootloader Where possible, enforcing secure boot. Tamper stickers on the device casing to indicate when a device has been opened
| Source | ID | Title |
|---|---|---|
EUVD |
EUVD-2022-40702 | The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network. |
No CVSS v4.0
No CVSS v3.0
No CVSS v2
This CVE is not in the KEV list.
Exploit prediction score: 1.0% (1.047% probability within 30 days)
| Link | Providers |
|---|---|
| https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01 |
|
| Type | Values Removed | Values Added |
|---|---|---|
| Metrics |
ssvc
|