CVE-2022-38069

Contec Health CMS8000

Published Sep 13, 2022 CVSS 3.1 · 4.3 Medium EPSS < 1% KEV · Not listed
Description
Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables the extraction of sensitive patient information or modification of device parameters

Remediation

Vendor Workaround

Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information. The following mitigations could assist in reducing the risk for exploitation of vulnerabilities: Disabling UART functionality at the CPU level Enforcing unique device authentication before granting access to the terminal / bootloader Where possible, enforcing secure boot. Tamper stickers on the device casing to indicate when a device has been opened

Tracking

Sign in to view the affected projects.

Advisories

Source ID Title
EUVD EUVD EUVD-2022-40674 Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables the extraction of sensitive patient information or modification of device parameters
History
Type Values Removed Values Added
Metrics ssvc

{'options': {'Automatable': 'no', 'Exploitation': 'none', 'Technical Impact': 'partial'}, 'version': '2.0.3'}

Weaknesses

Sources

Status
PUBLISHED
Assigner
icscert
Published
Updated
Apr 16, 2025, 4:10 PM
Reserved
Aug 29, 2022, 12:00 AM
Status
Modified
Published
Sep 13, 2022, 3:15 PM
Modified
Jun 17, 2026, 4:56 AM