CVE-2022-26388

Use of Hard-Coded Password Vulnerability in ELI Electrocardiograph Devices

Published Feb 7, 2025 CVSS 3.1 · 6.4 Medium EPSS < 1% KEV · Not listed
Description
A use of hard-coded password vulnerability may allow authentication abuse.This issue affects ELI 380 Resting Electrocardiograph:

Versions 2.6.0 and prior; ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph:

Versions 2.3.1 and prior; ELI 250c/BUR 250c Resting Electrocardiograph:

Versions 2.1.2 and prior; ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph:

Versions 2.2.0 and prior.

Remediation

Vendor Solution

Hillrom has released software updates for all impacted devices to address these vulnerabilities. New product versions that mitigate these vulnerabilities are available as follows: * Welch Allyn ELI 380 Resting Electrocardiograph: available Q4 2023 * Welch Allyn ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph: available May 2022 * Welch Allyn ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: available Q4 2023 Hillrom recommends users upgrade to the latest product versions. Information on how to update these products can be found on the Hillrom disclosure page https://hillrom.com/en/responsible-disclosures/ .


Vendor Workaround

Hillrom recommends the following workarounds to help reduce risk: * Apply proper network and physical security controls. * Ensure a unique encryption key is configured for ELI Link and Cardiograph. * Where possible, use a firewall to prevent communication on Port 21 FTP service, Port 22 SSH (Secure Shell Connection), and Port 23 Telnet service.

Tracking

Sign in to view the affected projects.

Advisories

Source ID Title
EUVD EUVD EUVD-2022-30947 A use of hard-coded password vulnerability may allow authentication abuse.This issue affects ELI 380 Resting Electrocardiograph: Versions 2.6.0 and prior; ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph: Versions 2.3.1 and prior; ELI 250c/BUR 250c Resting Electrocardiograph: Versions 2.1.2 and prior; ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: Versions 2.2.0 and prior.
History
Type Values Removed Values Added
Metrics ssvc

{'options': {'Automatable': 'no', 'Exploitation': 'none', 'Technical Impact': 'total'}, 'version': '2.0.3'}

Type Values Removed Values Added
Description A use of hard-coded password vulnerability may allow authentication abuse.This issue affects ELI 380 Resting Electrocardiograph: Versions 2.6.0 and prior; ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph: Versions 2.3.1 and prior; ELI 250c/BUR 250c Resting Electrocardiograph: Versions 2.1.2 and prior; ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: Versions 2.2.0 and prior.
Title Use of Hard-Coded Password Vulnerability in ELI Electrocardiograph Devices
Weaknesses CWE-259
References
Metrics cvssV3_1

{'score': 6.4, 'vector': 'CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:L'}

Subscriptions

No data.

Weaknesses

Sources

Status
PUBLISHED
Assigner
Baxter
Published
Updated
Feb 7, 2025, 6:50 PM
Reserved
Mar 3, 2022, 10:04 PM
Status
Deferred
Published
Feb 7, 2025, 5:15 PM
Modified
Jun 17, 2026, 4:35 AM