Use of Hard-Coded Password Vulnerability in ELI Electrocardiograph Devices
Default status is the baseline for the product, each version can override it (e.g. patched versions marked unaffected).
| Vendor | Product | Default status | Versions | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Welch Allyn | ELI 380 Resting Electrocardiograph | unaffected |
|
||||||
| Welch Allyn | ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph | unaffected |
|
||||||
| Welch Allyn | ELI 250c/BUR 250c Resting Electrocardiograph | unaffected |
|
||||||
| Welch Allyn | ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph | unaffected |
|
No data.
No data.
No data available yet.
Vendor Solution
Hillrom has released software updates for all impacted devices to address these vulnerabilities. New product versions that mitigate these vulnerabilities are available as follows: * Welch Allyn ELI 380 Resting Electrocardiograph: available Q4 2023 * Welch Allyn ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph: available May 2022 * Welch Allyn ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: available Q4 2023 Hillrom recommends users upgrade to the latest product versions. Information on how to update these products can be found on the Hillrom disclosure page https://hillrom.com/en/responsible-disclosures/ .
Vendor Workaround
Hillrom recommends the following workarounds to help reduce risk: * Apply proper network and physical security controls. * Ensure a unique encryption key is configured for ELI Link and Cardiograph. * Where possible, use a firewall to prevent communication on Port 21 FTP service, Port 22 SSH (Secure Shell Connection), and Port 23 Telnet service.
| Source | ID | Title |
|---|---|---|
EUVD |
EUVD-2022-30947 | A use of hard-coded password vulnerability may allow authentication abuse.This issue affects ELI 380 Resting Electrocardiograph: Versions 2.6.0 and prior; ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph: Versions 2.3.1 and prior; ELI 250c/BUR 250c Resting Electrocardiograph: Versions 2.1.2 and prior; ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: Versions 2.2.0 and prior. |
No CVSS v4.0
No CVSS v3.0
No CVSS v2
This CVE is not in the KEV list.
Exploit prediction score: < 1% (0.266% probability within 30 days)
| Type | Values Removed | Values Added |
|---|---|---|
| Metrics |
ssvc
|
| Type | Values Removed | Values Added |
|---|---|---|
| Description | A use of hard-coded password vulnerability may allow authentication abuse.This issue affects ELI 380 Resting Electrocardiograph: Versions 2.6.0 and prior; ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph: Versions 2.3.1 and prior; ELI 250c/BUR 250c Resting Electrocardiograph: Versions 2.1.2 and prior; ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: Versions 2.2.0 and prior. | |
| Title | Use of Hard-Coded Password Vulnerability in ELI Electrocardiograph Devices | |
| Weaknesses | CWE-259 | |
| References |
| |
| Metrics |
cvssV3_1
|
No data.