Improper Access Control for Boston Scientific Zoom Latitude
Default status is the baseline for the product, each version can override it (e.g. patched versions marked unaffected).
| Vendor | Product | Default status | Versions | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Boston Scientific | ZOOM LATITUDE | affected |
|
Configuration 1 [-]
| AND |
|
No data.
No data available yet.
Vendor Workaround
Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security, the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120.
| Source | ID | Title |
|---|---|---|
EUVD |
EUVD-2021-24844 | A skilled attacker with physical access to the affected device can gain access to the hard disk drive of the device to change the telemetry region and could use this setting to interrogate or program an implantable device in any region in the world. |
No CVSS v4.0
No CVSS v3.0
This CVE is not in the KEV list.
Exploit prediction score: < 1% (0.259% probability within 30 days)
Key SSVC decision points have not yet been added.
| Link | Providers |
|---|---|
| https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01 |
|
No history.