Dräger Infinity Explorer C700 Privilege Escalation via Kiosk Mode Bypass
Default status is the baseline for the product, each version can override it (e.g. patched versions marked unaffected).
| Vendor | Product | Default status | Versions | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Dräger | Infinity Explorer C700 | unaffected |
|
Configuration 1 [-]
| AND |
|
No data.
| Vendor | Product | Versions | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Draeger | Infinity Explorer C700 |
|
No CVSS v3.0
No CVSS v2
This CVE is not in the KEV list.
Exploit prediction score: < 1% (0.118% probability within 30 days)
| Type | Values Removed | Values Added |
|---|---|---|
| CPEs | cpe:2.3:h:draeger:infinity_explorer_c700:infinity_explorer_c700:*:*:*:*:*:*:* |
| Type | Values Removed | Values Added |
|---|---|---|
| References |
|
| Type | Values Removed | Values Added |
|---|---|---|
| Metrics |
ssvc
|
| Type | Values Removed | Values Added |
|---|---|---|
| First Time appeared |
Draeger
Draeger infinity Explorer C700 |
|
| Vendors & Products |
Draeger
Draeger infinity Explorer C700 |
| Type | Values Removed | Values Added |
|---|---|---|
| Description | Dräger Infinity Explorer C700 contains a privilege escalation vulnerability that allows attackers to break out of kiosk mode and access the underlying operating system through a specific dialog interaction. Attackers can exploit this kiosk escape to take control of the operating system and cause the device to display incorrect or no information from the connected Delta Family patient monitor. | |
| Title | Dräger Infinity Explorer C700 Privilege Escalation via Kiosk Mode Bypass | |
| Weaknesses | CWE-451 | |
| References |
| |
| Metrics |
cvssV3_1
|