CVE-2017-12725

Published Feb 15, 2018 CVSS 3.0 · 5.6 Medium EPSS < 1% KEV · Not listed
Description
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.

Remediation

No solution or workaround provided in the CVE record

Tracking

Sign in to view the affected projects.

Advisories

Source ID Title
EUVD EUVD EUVD-2017-4264 A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.
History

No history.

Weaknesses

Sources

Status
PUBLISHED
Assigner
icscert
Published
Updated
Aug 5, 2024, 6:43 PM
Reserved
Aug 9, 2017, 12:00 AM
Status
Modified
Published
Feb 15, 2018, 10:29 AM
Modified
Jun 17, 2026, 1:03 AM