Default status is the baseline for the product, each version can override it (e.g. patched versions marked unaffected).
| Vendor | Product | Default status | Versions | ||||||
|---|---|---|---|---|---|---|---|---|---|
| n/a | Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump | affected |
|
Configuration 1 [-]
| AND |
|
No data.
No data available yet.
| Source | ID | Title |
|---|---|---|
EUVD |
EUVD-2017-4263 | A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump contains hardcoded credentials, which are not fully initialized. The FTP server is only accessible if the pump is configured to allow FTP connections. |
No CVSS v4.0
No CVSS v3.1
This CVE is not in the KEV list.
Exploit prediction score: 1.3% (1.323% probability within 30 days)
Key SSVC decision points have not yet been added.