Default status is the baseline for the product, each version can override it (e.g. patched versions marked unaffected).
| Vendor | Product | Default status | Versions | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Hospira | LifeCare PCA Infusion System | affected |
|
Configuration 1 [-]
| AND |
|
No data.
No data available yet.
| Source | ID | Title |
|---|---|---|
EUVD |
EUVD-2015-1156 | Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access. |
No CVSS v4.0
No CVSS v3.1
This CVE is not in the KEV list.
Exploit prediction score: < 1% (0.771% probability within 30 days)
Key SSVC decision points have not yet been added.
| Link | Providers |
|---|---|
| https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01 |
|
No history.