Search vulnerabilities

Results 4

Export CSV
CVE-2023-30561

Lack of Cryptographic Security of IUI Bus

The data flowing between the PCU and its modules is insecure. A threat actor with physical access could potentially read or modify data by attaching a specially crafted device while an infusion is running.

CVSS 3.1 · 6.1 Medium EPSS < 1% KEV · Not listed
Vendors
Bd
CVE-2020-25165

BD Alaris PC Unit, Model 8015, Versions 9.33.1 and earlier and BD Alaris Systems Manager, Versions 4.33 and earlier The affected products are vulnerable to a network session authentication vulnerability within the authentication process between specified versions of the BD Alaris PC Unit and the BD Alaris Systems Manager. If exploited, an attacker could perform a denial-of-service attack on the BD Alaris PC Unit by modifying the configuration headers of data in transit. A denial-of-service attack could lead to a drop in the wireless capability of the BD Alaris PC Unit, resulting in manual operation of the PC Unit.

CVSS 3.1 · 7.5 High EPSS 1.7% KEV · Not listed
Vendors
Bd